E1070 Service Manual 133
4.3 Enable/Configure Weigh-Bars
4.3 Enable/Configure Weigh-Bars
1. Access the Service (0701) menu…
CAL is displayed.
2. Press the TARE key…
oPtion is displayed.
3. Press the PRINT key…
A-Out is displayed.
4. Repeatedly press the UNITS key until SEnSor is displayed.
5. Press the PRINT key…
gHoSt is displayed.
6. Press the TARE key
CELLS is displayed.
7. Press PRINT
1 is displayed.
8. Press PRINT
on or oFF is displayed.
9. Toggle between the choices by pressing the UNITS or TARE key. Press
ENTER to accept when on is displayed.
10. Press UNITS to advance to the next cell.
11. Repeat steps 7-10 for the correct number of cells being connected.