16 E1070 Service Manual
2 Introduction
2.2 Error Messages
The following are displayed messages you may see if problems occur or if invalid
operations are attempted with your indicator:
When you are in the Linearity menu item in the Service menu, you may see the
following errors:.
Display Description
Overrange weight. Scale is overloaded.
Underrange weight. Scale is underloaded.
The unit cannot perform a function. Displayed only while key is held down.
Displayed while a key is pressed when attempting to modify a sealed selection
without edit privileges.
Display Description
Out of ascending order.
Entered value is less than 1% of scale capacity.
Entered value causes resolution of greater than 100,000 divisions.
Linear points must be done in order from lightest weight to heaviestWhen you are in
the Span menu item in the Service menu, you may see the following errors: