E1070 Service Manual 29
3.4 Service Menu
Source (Analog or SensorComm)
1. From the SCALE display, press the PRINT key…
SourCE is displayed. Use this item to choose between an analog or
SensorComm based system.
2. Press the PRINT key…
The current setting is displayed.
3. Toggle between the Analog and Scom (North America only) choices using the
TARE or UNITS key. When your choice is displayed press the ENTER key…
SourCE is displayed.
CAP. (Capacity)
1. With SourCE displayed, press the UNITS key…
CAP. is displayed. Use this item to set the capacity for the scale.
2. Press the PRINT key…
The current capacity value is shown.
3. Press ENTER to accept this value or key in a new capacity and press
ENTER…
CAP. is displayed.
DIV. (Division)
This item and the next one, DP.POS., set the division size.
1. With CAP. displayed, press the UNITS key…
div. is displayed. This stands for the division value of your displayed
weight.
2. Press the PRINT key…
The current division value is shown.
3. Scroll through the choices by using the TARE or UNITS key.
Pick from the following values; 1, 2, 5, 10, 20, 50, 100, 200, 500, 1/2, 2/
5, 5/10, 10/20, 20/50.
The fraction choices (in bold) are for use as dual range divisions.
The first number is the division value for the first half of the
capacity and the second number is the division value for the 2nd
half of the capacity.
All of these capacities function in conjunction with the decimal place
position. For example, if you choose a division value of 5 and a decimal
position of 12345.6, your division size will be .5.
4. When your choice is displayed, press ENTER.
div. is displayed.