Model E1070 Indicator Service Manual
OPTION 29. Press the UNITS key. . .
OPTION is displayed.
30. Press the PRINT key. . .
A-OUT is displayed. This stands for the analog output test.
31. Press the PRINT key. . .
A numeric entry screen is displayed.
32. Key in a percentage between 0 and 100 and press the ENTER key. . .
The analog output will put out that percentage of voltage. For
example: If you have output set from 0 to 10V and you key in a
percentage of 25, the analog output voltage should read 2.5 volts.
You can continue to key in other percentages, press ENTER and
check the analog output voltage.
33. When you are finished testing the analog output, press the ESC key. . .
A-OUT is displayed.
34. Press the UNITS key. . .
CNTR is displayed. This checks the pulse counter.
35. Press the PRINT key to check the counter. . .
0 is displayed if the pulse counter is inactive. If pulse input are
occurring the display will show the increasing pulse counts.
36. Press SELECT or ENTER to stop the test. . .
CNTR is displayed.
37. Press the UNITS key. . .
SENSOR is displayed. Use this to test the function of each weight
sensor attached via SensorComm.
38. Press the PRINT key. . .
1 is displayed.
39. Scroll through the available weight sensor numbers by pressing the
TARE or UNITS key. Press the ENTER key when the sensor you want
to test is displayed. . .
mV/V output of the selected sensor is displayed.
40. Apply weight to the scale to verify the mV/V level changes.
41. Press the ENTER key to exit the test. . .
Sensor number is displayed.
42. Repeat steps 38 through 40 for each sensor you want to test.
43. Press the SELECT key. . .
SENSOR is displayed.
A-OUT test
CNTR test
SENSOR test
This test can be used to
properly adjust the deadload of
the scale. Each sensor should
have similar mV/V outputs if
the physical load on the scale
is equally distributed.
Supervisor Menu—(continued)