Model E1070 Indicator Service Manual
CAP.
(Capacity)
1. From previous step 5, press the UNITS key. . .
CAP. is displayed. Use this item to set the capacity for the scale.
2. Press the PRINT key. . .
The current capacity value is shown.
3. Press ENTER to accept this value or key in a new capacity and press
ENTER. . .
CAP. is displayed.
This item and the next one, DP.POS., set the division size.
1. From previous step 3, press the UNITS key. . .
DIV. is displayed. This stands for the division value of your
displayed weight.
2. Press the PRINT key. . .
The current division value is shown.
3. Scroll through the choices by using the TARE or UNITS key.
Pick from the following values; 1, 2, 5, 10, 20, 50, 100, 200, 500,
1/2, 2/5, 5/10, 10/20,
20/50.
The fraction choices are for use as dual range divisions. The
first number is the division value for the first half of the
capacity and the second number is the division value for the
2nd half of the capacity.
All of these capacities function in conjunction with the decimal
place position. For example, if you choose a division value of 5
and a decimal position of 12345.6, your division size will be .5.
4. When your choice is displayed, press ENTER.
DIV. is displayed.
Use this item to set the decimal point position in the displayed weight.
1. From previous step 4, press the UNITS key. . .
DP.POS. is displayed. This stands for decimal point position.
2. Press the PRINT key. . .
The current decimal point position is shown. Choices available
are; 123456, 12345.6, 1234.56, 123.456, 12.3456 and 1.23456.
3. Scroll through the choices by using the TARE or UNITS key. When your
choice is displayed, press ENTER.
DP.POS. is displayed.
DIV.
(Division)
Service Menu—SCALE submenu (continued)
DP.POS.
(Decimal point position)