Page #13 background image
Power connections Page 11/51
MID_M2000_BA_02_1807
BA17MID
weiß (45)
braun (46)
pink (40)
schwarz (44)
schwarz (44)
gelb/grün (13)
gelb (12)
grün (11)
5 bis 300 m
grün (11)
gelb (12)
pink
(40)
weiß (45)
braun (46)
schwarz (44)
schwarz (44)
gelb/grün (13)
5 to 150 m
pink
4.2.1 Signal cable specification
Note: Only use signal cables delivered by Badger Meter or corresponding
cable in accordance with the following specification.
Take max. signal cable length between detector and amplifier into
account (keep distance as low as possible).
Distance
With electrode idle
Loop resistance
0 75 m
3 x (2 x 0,25 mm²)
=< 160 /km
> 75 150 m
3 x (2 x 0,50 mm²)
=< 80
/km
PVC cable with pair and total shield
Capacity: wire/wire < 120 nF/km, wire/shield < 160 nF/km
Temperature range 30 to +70 °C
Maximum cable length at different fluid temperatures
0
25
50
75
100
125
150
175
200
225
250
275
300
0 25 50 75 100 125 150 175
Cable length [m]
Temperature [°C]
0,25 mm²
0,5 mm²
0,75 mm²
BA17MID
white
white
brown
brown
black
black
black
black
yellow/green (13)
yellow/green (13)
yellow (12)
yellow (12)
green (11)
green (11)

Other manuals for Badger Meter ModMAG M2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Badger Meter ModMAG M2000 and is the answer not in the manual?

Badger Meter ModMAG M2000 Specifications

General IconGeneral
BrandBadger Meter
ModelModMAG M2000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals