10-57
Sensitivity = 1
µ
A/V 100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Specific Parameters
Real Time Integrator = Off Off/On
Quiet Time (sec) = 2 0 to 65535
Comments
a) The
Sample Interval
is the time resolution of the experiment.
b) The number of data points recorded in an experiment must be less than or
equal to 8000, which is the data storage limit. If the product exceeds 8000,
then the
Run Time
is automatically adjusted to an acceptable value.
c) A
Precondition
potential can be applied for the
Precondition Time
before
the start of the experiment.
d) When enabled, the
Real Time Integrator
adds all the current values to give
the charge accumulated.
Graphics Menu
Single Graph
displays the current vs. time plot.
Analysis Menu
Results Graph
is disabled for
TB
.
Mode = DPTB
General Parameters
Initial E (mV) = 0 -3276 to 3276
Final E (mV) = 0 -3276 to 3276
Pulse Width (msec) = 250 25 to 20000
Pulse Period (msec) = 500 50 to 40000
Sample Width (msec) = 200 1 to 19995
Run Time (sec) = 4000 1 to 65535
Sensitivity = 1
µ
A/V 100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Specific Parameters
Quiet Time (sec) = 2 0 to 65535