10-58
Comments
a)
Pulse Width
is the duration of the potential pulse.
b)
Sample Width
is the time at the end of the pulse during which the current is
measured. The maximum value is
Pulse Width
- 5 msec. The current is
sampled 16 times per msec.
c)
Pulse Period
is the time for one potential cycle, and must be at least twice the
Pulse Width
.
d) The number of data points recorded in an experiment must be less than of
equal to 8000, which is the data storage limit. If the product exceeds 8000,
then the
Run Time
is automatically adjusted to an acceptable value.
Graphics Menu
Single Graph
displays the differential current vs. time plot.
Analysis Menu
Results Graph
is disabled for
DPTB
.
Mode = TPTB
General Parameters
Initial E (mV) = 0 -3276 to 3276
Display Data Set = 3 1, 2 or 3
E (mv) = 0 -3276 to 3276
Pulse Width (msec) = 50 10 to 65535
Sample Width (msec) = 17 1 to 32767
Increment E (mV) = 0 -100 to 100
Sensitivity = 1
µ
A/V 100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Number of Cycles = 1 1 to 2000
Specific Parameters
Quiet Time (sec) = 2 0 to 65535
Comments
a)
E
is the amplitude of the potential pulse.
b)
Pulse Width
is the duration of the potential pulse