10-26
Graphics Menu
Single Graph
displays the current vs. potential plot.
Analysis Menu
The
Auto
option for
Results Graph
displays the current vs. potential plot,
and the half-wave potential and limiting current are listed in the
Main
window. Alternative baselines can be set by the user through the
Manual
option.
Mode = NPV/P
General Parameters
Initial E (mV) = 0 -3276 to 3276
Final E (mV) = 0 -3276 to 3276
Sensitivity = 1
µ
A/V 100 mA/V to 100 nA/V (10 pA/V with the
Low Current Module)
Specific Parameters
Scan Rate (mV/sec) = 20 (4) 1 to 200
Sample Width (msec) = 17 1 to 250
Pulse Width (msec) = 50 3 to 2000
Pulse Period (msec) = 200 40 to 8000
Drop Time (msec) = 1000 100 to 8000
Quiet Time (sec) = 2 (0) 0 to 65535
Comments
a) The
Pulse Period
is used in
NPV
, the
Drop Time
in
NPP
. When thedefault
parameter value for
NPP
is different from that for
NPV
, it is shown in
parentheses
b) Increasing the
Sample Width
can reduce the response noise.
c) The sample interval is equal to the product of the
Scan Rate
and the
Pulse
Period
(or
Drop Time
) (e.g., the default value is 4 mV).