ADVANCED FUNCTIONS OF AK 98
Sodium, Bicarbonate Profiling
1. Press the Fluid button.
2. Select the UF rate tab or Cond tab.
3. Press Profiling and press Sodium (Na
+
) or Bicarbonate (HCO
3
-
) to set
up profiling.
4. Set up the Start and Stop values.
5. Press Activate.
Note: You can create and use models by having an authorized
technician preset the values for each profile.
The sodium (Na
+
) or bicarbonate (HCO
3
-
) profiles may consist of increasing or decreasing concentrations in the dialysate.
Note: If the treatment time changes during sodium and/or bicarbonate profiling, the profiling graph will not
change. Therefore, even if the treatment time decreases, it will not reach the preset stop value. If sodium
or bicarbonate profiling is deactivated during treatment, the machine will continue to run using the
values from the point when profiling stopped. If profiling is reactivated without a change to the profiling
parameters, the machine will continue to run from the point when profiling was deactivated.
Example of linear graphs for sodium and bicarbonate profiles
Time
START
Concentration
Increase Decrease
Time
START
STOP
STOP
Concentration
Profiling

Table of Contents

Other manuals for baxter AK 98

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter AK 98 and is the answer not in the manual?

baxter AK 98 Specifications

General IconGeneral
Brandbaxter
ModelAK 98
CategoryMedical Equipment
LanguageEnglish

Related product manuals