Chapter 4: Maintenance 67
Decontaminate the
fluidics system for
storage
Cleans out the fluidics lines with
BD FACSClean solution, then
fills them with
BD FACS shutdown solution
See Decontaminating the
Fluidics System for Storage on
page 80.
Before long-term storage
Replace the Bal seal Replaces worn Bal seal
See Replacing the Bal Seal on
page 80.
When the seal becomes
worn or cracked (sample
tubes will not pressurize)
Reset the cytometer
circuit breaker
See Resetting the Cytometer
Circuit Breaker on page 84.
After the circuit breaker
is tripped
Reattach Ethernet and
RS232 cables
See Reconnecting the Ethernet
and RS232 Cables on page 85.
When a cable is
disconnected
Reconnect fluidics cart
tubing
See Reconnecting the Fluidics
Cart Tubing on page 85.
As needed
Replace fluidics level
sensors
See Replacing the Fluidics Level
Sensors on page 88.
When instructed to by a
BD Biosciences service
representative
Replace fluidics cart fuses See Replacing the Fluidics Cart
Fuses on page 91.
When cart will not
function because of power
surge or other electrical
event
Procedure Description and Location Frequency