EasyManua.ls Logo

BD FACSCanto - Analysis View and Data Analysis

BD FACSCanto
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48 BD FACSCanto Flow Cytometer Reference Manual
5 Aim the barcode reader at the center of the barcode. The barcode reader
has two focal lengths, 4 and 9 inches:
• For reading BD FACS 7-color setup bead labels, hold the barcode
reader 9 inches from the labels.
• For reading patient information from sample tube labels, hold the
barcode reader 4 inches from the labels.
6 Press and hold the trigger on the barcode reader until you hear a beep.
If the reader does not beep, adjust your distance from the barcode while
continuing to hold the trigger.
; Tip To obtain a reading, keep the bar code reader aimed at the center of
the label. Do NOT sweep across the label.
7 Compare software field values with the setup beads or sample label.
Cleaning the Barcode Reader
For best performance, keep the front window of the barcode reader clean. Do not
touch the window directly, and wipe it only with a soft, non-abrasive cloth
moistened with one of the following:
• isopropyl alcohol
• ethyl alcohol (denatured grade)
To prevent laser injury, do not stare into the reader or aim it at another
person’s eyes while the trigger is depressed.
Do not use BD FACSClean solution (or bleach) to clean or disinfect the
barcode reader.

Table of Contents

Other manuals for BD FACSCanto

Related product manuals