Chapter 4: Maintenance 91
Replacing the Fluidics Cart Fuses
Power surges and other electrical events could cause a fuse to blow. Use the
following procedure to replace the fuses.
Materials
• two replacement fuses (see Installation Kit on page 128 for ordering)
• small screwdriver
Removing the Old Fuses
1 Shut down the cytometer and turn off the cytometer power.
2 Unplug the power cord from the wall outlet and the fluidics cart.
Removing the plug allows easier access to the fuse door.
3 Insert a small screwdriver into the slot and gently pry the fuse door open.
To protect against shock, always turn off the cytometer and unplug the
power cord before performing this procedure.
slot for screwdriver
empty socket,
plug removed
fuse access door
circuit breaker