Chapter 4: Maintenance 73
3 Install bypass tubing in place of the filter.
Push the tubing into each quick-disconnect port until you hear a click.
4 Select Cytometer > Cleaning Modes > Prime After Tank Refill.
5 Select the checkbox that corresponds to the filter you have bypassed.
6 Click OK.
7 When the prime finishes, remove the bypass tubing.
8 Reattach the filter to the fluidics cart.
9 Repeat the Prime After Tank Refill command.
bypass tubing,
installed
BD FACSDiva software
BD FACSCanto clinical software