EasyManua.ls Logo

Benning IT 110 - Page 31

Benning IT 110
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
Fig. 5.14: Contact voltage measurement menu
Connect the test cable to the BENNING IT 110/ BENNING IT 120 B.
Step 2 Set the following measuring parameters and limit values:
- Nominal residual current
- RCD type
- Limit contact voltaget
Step 3 Followtheconnectiondiagramshowningure5.15toperformcontactvoltagemeasu-
rement. Use the Help function if necessary
L1
L2
L3
N
PE
R
E
Ro
LPEN
L
/
L
1
N
/
L
2
P
E
/L
3
Fig. 5.15: Connection of plug test cable or universal test cable
Step 4 Check the displayed warnings and online voltage/terminal monitor before starting the
measurement. If OK, press the TEST key. After performing, the measurement results
with PASS/FAIL indication appear on the display.
Fig. 5.16: Example of contact voltage measurement results
Displayed results:
U: Contact voltage
Rl: Loop impedance (fault loop resistance)
Save displayed results for documentation purposes. Refer to chapter 6.1 for further information on
setting functions and saving of measurement results. (BENNING IT 120 B only)
Note:
- Parameters set in this function are also kept for other RCD functions.
- The measurement of contact voltage does not normally trip an RCD. However, the trip limit may
beexceededasaresultofleakagecurrentowingtothePEprotectiveconductororviaaca-
pacitive connection between L and PE conductors.

Table of Contents

Related product manuals