01/10 347 LB 444 K-40
23
6.4 Product Data
Enter product data for parameter setup and calibration separately for each product.
Product selection
No. 1/2/3/4:
Enter product number for calibration.
Current Output:
Select current output signal 0 or 4 mA.
Current Output Limits:
0/4 mA:
20 mA:
Example:
Measuring range 10 ... 30% K20.
Input: at 4 mA: 10
at 20 mA: 30
Current Output Error:
Current:
VALUE
In case of error, the current output jumps to the defined value ( 0...22 mA)
HOLD
In case of error, the current output holds the last value.
Caution: If you have selected “continue in case of error”, the current output
operates using a wrong value corresponding to the false value.
Relay No. 2 setup:
Minimum / Maximum / Det. Temp. / Mass pulse:
Different functions can be assigned to the relay. As soon as the status to be signaled
Concentration < minimum or
Concentration > minimum
Detector temperature > X °C
has been reached, the relay is de-energized. If the displayed value drops again, the relay
picks up again, delayed by the value of the selected hysteresis.
Hysteresis
The switching hysteresis should be about 5%.
Relay No. 3 setup:
Same as relay no. 2.
Minute printout
Value
The measured value can be printed out in defined time intervals using a printer
connected to the RS 232 interface.