01/10 347 LB 444 K-40
41
DC: Max. 300V, max. 1A, max. 60 W non-reactive load
Current input: 0/4 - 20 mA isolated
for input of temperature signal or
volume current signals
Interfaces: RS 232. Port for data transfer from evaluation unit to printer
or PC
8.2 Detectors
Type
Crystal ∅
/ l in mm
Weight
kg
Detector code Ex protection
LB 5401-03 50/50 18 0 No
LB 5430 150/150 52 23 No
Protection type: IP 65,
Housing: Stainless steel housing
Cable entry: M16 external cable diameters 7...10 mm for connection cable
detector - evaluation unit and M 12 for external cable diameter
7...10 for connection of Pt 100
Temperature range:
Operating
temperature:
- 20 to + 50°C. A water cooling device is available for higher
temperatures. Monitoring of detector temperature and alarm if
max. permissible temperature is exceeded.
Storage
temperature:
- -40 to + 70 °C
Connection cable evaluation unit - detector
Cross-section in mm² Max. cable length in m
1 1000
1.5 1500
2.5 2500