01/10 347 LB 444 K-40
36
7.8 Error Messages
7.8.1 Acknowledging Error Messages
All error messages must be reset with <enter>.
Several simultaneously or consecutively occurring errors are stored in an error register in
the order of their appearance; they have to be reset individually by pushing <enter>
several times.
All errors are signaled via the “Error” output (relay 16a/16c).
7.8.2 Error Messages during Operation
Error
no.
Message Cause, Remarks Remedy
32 Wrong
password
Wrong password entered to unlock
the keyboard.
Push “enter” to reset error
message. Enter correct
password.
7.8.3 Error Messages during Calibration
Error
no.
Message Cause, Remarks Remedy
10 Not calibrated Calibration incomplete Perform calibration
36 Missing calib.
points
Not enough points entered for
selected calibration mode
Enter further calibration
points or select another
calibration mode
37 Curve not clear System could not calculate a
clearly rising or falling curve.
Check calibration points for
input errors.
Select calibration mode
“linear”.
38 Error one-point
calibration
Absorption coefficient = 0 Absorption coefficient = 0
39 Measuring path
not defined
Measuring path = 0 Enter measuring path “1”
56 Range overflow Measuring path = 0 Enter measuring path “1”