01/10 347 LB 444 K-40
43
50 I.Max/I.Min value Current output entered incorrectly
51 Current input error Current input range entered incorrectly
52 Current output limit value Current output range entered incorrectly
53 Current input error Current input too high > 20mA
54 Device not calibrated Calibration has to be terminated with <enter>
56 Range overflow Input value outside permissible range
57 Wrong input value Input value outside permissible range
58 Calibration faulty Wrong calibration
60 Digital inputs for product
selection allowed
Window for product selection in menu Parameter set to
external.
Problem Possible Cause Potential Solution
No display No power supply Check mains cables
Check fuses
Display unreadable Processor error Note error code
Perform total reset
Replace evaluation unit
No pulse rate
(Error code 2)
Power supply for
detector interrupted
Check cable to detector
Detector is faulty Replace detector
Result reading
shows drifts
Detector stabilization
faulty
Replace detector
Photomultiplier faulty Replace photomultiplier
Decay compensation
activated