MK / MKT (E5) 03/2019 page 120/158
Fault description
Possible cause
Required measures
Controller (continued)
Program keeps the last program
setpoint constant while in setting
“ramp”.
Program line with setting
“ramp” is incomplete.
When programming with setting
“ramp”, define the end value of the
desired cycle by adding an additional
section with a section time of at least
one second.
Ramp temperature transitions are
only realized as steps.
Setting “step” has been select-
ed.
Select setting “ramp”.
Acknowledging the alarm does
not cancel the alarm state.
Cause of alarm persists.
Remove cause of alarm. If the alarm
state continues, contact BINDER
service.
Alarm message: “- - - - “ or
“<-<-<“ or “>->->“
Sensor rupture between sen-
sor and controller or Pt 100
sensor defective.
Contact BINDER service.
Short-circuit.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired chambers must comply with the BINDER quality standards.