Fluorescence Tests | 123
Protocol Name: Synergy HTX FI_B.prt
Parameter Default Setting
Plate Type: Greiner SensoPlate (Mfr. #655892)
Kinetic Loop: Run time :45 seconds
Interval: 3 seconds
Read Step Label: “Sensitivity Read”
Read Well: D7
Filter Set: EX 485/20 nm
EM 528/20 nm
Optics Position: Bottom
Gain: Auto
Scale to High Wells
Scale Well: D7
Scale Value: 30000
Delay after Plate Movement: 100 msec
Measurements per Data Point: 50
Dynamic Range: Standard
Light Source: Tungsten
Read Step 2
Kinetic Loop: Run time: 1 minute, 35 seconds
Interval: 6 seconds
Read Step Label: “Sensitivity Read Buffer”
Read Wells: C9–E9
Filter Set: EX 485/20 nm
EM 528/20 nm
Optics Position: Bottom
Gain: Auto
Use first filter set gain from FIRST
Read Step
Delay after Plate Movement: 100 msec
Measurements per Data Point: 50
Dynamic Range: Standard
Light Source: Tungsten
Read Step 3
Read Step Label: “Corners Read”
Synergy HTX Operator's Manual