124 | Chapter 6: Instrument Qualification
Protocol Name: Synergy HTX FI_B.prt
Parameter Default Setting
Read Wells: A1–A3, A10–A12, H1–H3, H10–H12
Filter Set: EX 485/20 nm
EM 528/20 nm
Optics Position: Bottom
Gain: Auto
Scale to High Wells
Scale Well: A3
Scale Value: 30000
Delay after Plate Movement: 100 msec
Measurements per Data Point: 50
Dynamic Range: Standard
Light Source: Tungsten
Read Step 4
Read Step Label: “Linearity Read”
Read Wells: C1–F5
Filter Set: EX 485/20 nm
EM 528/20 nm
Optics Position: Bottom
Gain: Auto
Scale to High Wells
Scale Well: C1
Scale Value: 30000
Delay after Plate Movement: 100 msec
Measurements per Data Point: 50
Dynamic Range: Standard
Light Source: Tungsten
Protocol Name: Synergy HTX FI_T.prt
Parameter Default Setting
Plate Type: Costar 96 black opaque, Mfr. #3915
Kinetic Loop: Run time :45 seconds
Interval: 3 seconds
Read Step Label: “Sensitivity Read”
BioTek Instruments, Inc.