By turning the knob, following options will be displayed:
By pressing Remote / Local key once again, the displayed interface is selected
and the specific parameters can be set (for RS 232 interface, baud rate and
parity will be set, for GPIB interface GPIB address of the power supply will be
set).
The parameter values can be changed by turning the knob. When desired value
for the parameter is displayed, Remote / Local key must be pressed.
Important note!
If you enter Remote / Local submenu and no action takes place for approx. 20
seconds, the power supply will leave this submenu. No Change message is
displayed and the power supply returns to the previous state (the state before
entering this submenu).
Important note!
If you enter Remote / Local submenu and no action takes place for approx. 20
seconds, the power supply will leave this submenu. No Change message is
displayed and the power supply returns to the previous state (the state before
entering this submenu).
Important note!
While in local mode, if by pressing Remote / Local key
message will be displayed, it means that the remote interface parameters cannot
be read from the non-volatile memory. The power supply must be turned off
and then turned on.
If this message is displayed again after pressing Remote / Local key in local
mode, the power supply must be delivered to B&K Precision for service.
RS - 232
GPIB / 488
I / O Error