EasyManuals Logo

BK Precision 9120 User Manual

BK Precision 9120
119 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
95
Standard Event Register
Standard Event Register reports different instrument events, such as: query
errors, device dependent error, execution errors, command errors, power on
event. A set bit indicates that an event of the specified type occurred.
Bit 0 - OPC (Operation Complete)
- indicates that all previous commands have been executed and the
device is ready to accpet new commands
- this bit is set only after an *OPC command
Bit 1 - Not used
- always set to 0
Bit 2 - QYE (Query Error)
Bit 3 - DDE (Device Dependent Error)
- indicates that a device dependent (user defined) error has occurred,
such as self test errors, or calibration errors
Bit 4 - EXE (Execution Error)
- indicates that an execution error occured
Bit 5 - CME (Command Error )
- indicates that a command error occured
Bit 6 - Not used
- always set to 0
Bit 7 - PON (Power On)
-
-
indicates that the power supply was turned off and on since the last time
Standard Event Register was cleared or read.
Standard Event Register is cleared by:
-
-
*CLS command
-
-
quering event register, using *ESR? Command
Standard Event Enable Register is cleared by:
-
-
*ESE 0 command