4.5.1 Country-specific settings
Parameter settings
Parameter Germany Standard EU Denmark Switz. France Holland Italy
Lambda calculation ON ON ON ON ON ON ON
CO
cor
calculation OFF OFF ON OFF ON OFF OFF
O
2
measurement ON ON ON ON ON ON ON
Adjustment data settings
Setting Germany Standard EU Denmark Switz. France Holland Italy
Recalibration interval 365 days 365 days 365 days 365 days 365 days 365 days 365 days
Gases for recalibration HC, CO, HC, CO, HC, CO, HC, CO, HC, CO, No setting HC, CO,
CO
2
CO
2
CO
2
CO
2
CO
2
CO
2
Reaction when recalibration Disable No setting No setting Disable No setting Message No setting
due
Leak test configuration Days Days NO Days NO Days Days
Leak test interval 1 day 1 day NO 1 day NO 1 day 1 day
HC residue test YES YES YES YES YES YES YES
National regul. complied with YES NO YES YES NO YES YES
Change CO
cor
setting Disabled Enabled Disabled Disabled Enabled Disabled Enabled
Service interval 183 days 183 days 183 days 183 days 183 days 183 days 183 days