Servicing the Unit
4
Infusomat® Space 6.0 4 - 3EN
Procedure:
1. Set the pump down with the rear facing upward; fix it in po-
sition, if necessary.
2. Fill the connector with contact cleaner. Depending on the ex-
tent of soiling, allow to work.
3. Loosen any incrustations wi
th a cleaning brush.
4. Moisten swabs with contact cleaner and clean the contacts
w
ith them.
5. Rinse connector with contact cleaner, dry off with com-
pressed air.
6. Perform a visual check, repeat cleaning, if necessary.
7. Spray cleaned contacts with contact spray; allow to evapo-
rate for several minutes.
8. Check that the connectors function correctly (connectio
n
problems remedied?).
Servicing the Battery
The instructions for use contain a detailed description on how to
service the battery.
If a battery module is not discharged co
mpletely for more than 28
days, a servicing program for the battery module can be started
on the unit.
Fig.: 4 - 4 Example: connector on SpaceStation