Infusomat® Space 6.0 5 - 1EN
Index h
(Master – to be added to the documentation)
M686 70 01 05 F04 / 38918838 Sheet 1 of 3
5 - Technical Safety Check (TSC)
Infusomat® Space
Checklist for Technical Safety Check – Every 24 Months User
Unit: Infusion Pump Infusomat® Space
Manufacturer: B. Braun Melsungen AG
Observe the Service Manual and the instructions for use. All measured values are to be
documented. Accessories used should be included in testing. Make exclusive use of cali-
brated measuring equipment.
Year of Procurement
Article No. Unit No. Stock No.
Visual Inspection Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection
Infusomat® Space:
Cleanliness, completeness, damage and
defects affecting safety, damage and
readability of the labels.
Particularly:
Connectors „P2“ and „P3“
Safety clamp
Pressure sensor, downstream
(membrane)
Pressure sensor, upstream (mem-
brane)
Pump membrane
Pump slide
Coding of pump slide guide
Pump connecting rod
Air inline sensor (free from grease
or ultrasonic gel, free from cracks)
Drop sensor (lens)
Accessories:
Cleanliness, completeness, damage and
faults affecting safety, damage and
readability of the label.
Check the unit and the accessories for
compatibility
The patient and housing leakage current of
the Infusomat® Space is caused exclusively
by the operating voltage supply (Power Sup-
ply SP or SpaceStation).
The Technical Safety Checks of the power
supply SP (drawing No. M001321005F04) or
of the SpaceStation (drawing No.
M690000046F04) serve to check whether
both limit values are met.
Locking with second unit
Battery compartment cover
Determine age of battery
If older than 8 years: recommend re-
placement of battery or perform bat-
tery check (see “Procedural
Instructions on the TSC)
Switch on unit with power supply:
Self-test
Indicator lamps (LEDs)
Audible alarm
Visual alarm
Status display
Lighting
Operation
Opening and closing mechanism of the
operating unit
Slide clamp releases release lever of the
safety clamp. The Infusomat® Space
Line is squeezed.
Buttons on the operating unit
Infusion
Staff call
5 - 4
Edit