5 - Technical Safety Check (TSC)
5 - 2 Infusomat® Space 6.0 EN
Index h
(Master – to be added to the documentation)
M686 70 01 05 F04 / 38918838 Sheet 2 of 3
Operation (continued)
Trigger bolus at the device
Trigger bolus by pressing the PCA but-
ton
Slide clamp is completely pulled on to
the Infusomat® Space Line when the
operating unit is opened.
The operating lever of the safety clamp
can be pressed down uniformly against
spring tension and engages. The yellow
caution sign flashes.
Pressure cut-off, electronic:
Delivery rate: 250 ml/h
Test without drop sensor
Upstream
Alarm
Downstream
Pressure stage 2
(0,1 … 0,7 bar) ____ bar
Pressure stage 5
(0,4 … 1,0 bar ____ bar
Pressure stage 9
(0,8 … 1,4 bar) ____ bar
Pressure limitation, mechanical:
Pmax (1,8 … 2,5 bar) ____ bar
Pmin (> 1,5 bar) ____ bar
Safety clamp
Pmin (> 1 bar) ____ bar
Delivery accuracy
Ambient temperature 20 … 25°C,
Delivery rate: 250 ml/h, Measured volume: 25 ml
Divergence (± 5 %) ____ %
Drop sensor (optional)
Delivery rate: 250 ml/h
Drop sensor alarm, flow
Drop sensor alarm, occlusion
Visual Inspection Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection