5 - Technical Safety Check (TSC)
Infusomat® Space 6.0 5 - 3EN
M686 70 01 05 F04 / 38918838 Sheet 3 of 3
Index h
(Master – to be added to the documentation)
Air inline sensor
Water value (> 600 mV) ____ mV
Difference of temperature
sensors (< 3 °C) ____ °C
Air alarm
Air value (< 100 mV) ____ mV
Note: If the water or air value is not
reached immediately, these values must be
read again after 120 seconds.
Switch on unit without power supply
Self-test
Magnetic function of the battery com-
partment cover
Remove battery module
Alarm for min. 3 minutes
Mech. Aids and Measuring Equip. Used Accessories Used
Disposables
Type: _______________________
Part No.: __________________
Manometer
Ident. No.: ____________________
Calibrated until ________________
Scales
Ident. No.: ____________________
Calibrated until ________________
Graduated cylinder
Stop watch
Service connector SP
__________________________
__________________________
Staff call line
Power Supply SP
Battery module
Space PCA Kit (PCA button)
Infusomat® Space Line, standard type
Drop sensor
__________________________
__________________________
Test result: Defects found which could endanger patients, users or
third parties: Yes No
Inspection performed by:
Measures to be taken: None Repair
Unit handed over on:
Replacement of battery recommended
__________________________________________
To:
Special features / documentation:
___________________________________________________________________________
Date / Signature:
___________________________________________________________________________ Next deadline:
Visual Inspection Electrical Safety
According to IEC 60601-1 or IEC 62353
Functional Inspection