Procedural Instructions on the TSC
Infusomat® Space 6.0 6 - 7EN
4. Vent test setup, position of the three-way cocks please see
Fig.: 6 - 6, start infusion and deliver first of all in to an open
system (without pressure gauge).
Upstream
5. Push roller clamp of the Infuso
mat® Space Line up to approx.
30 cm before the pump and close the roller clamp or pinch
the Infusomat® Space Line between the container and 30 cm
before the pump. An alarm is output.
6. Open roller clamp.
Downstream
7. Switch over three-way valve, please see Fig.: 6 - 7, and deliv-
er towards the pressure gauge.
Read off maximum value on the p
ressure gauge upon an
alarm and before an automatic pressure reduction and com-
pare with the specifications in the TSC.
8. Check all pressure stages listed in the TSC and document val-
ues.
Pressure Limitation, Mechanical
The optional drop sensor must not be connected.
Do not use an Infusomat® Space Line with an upstream "Y" site.
The back check-valve associated with the upstream "Y" site may
skew Mechanical Pressure Limitation test results.
P
max
and P
min
1. Plug service connector Space on connector “P2” so that the
ele
ctronic pressure cut-off is deactivated.
2. Input a delivery rate according to t
he TSC, start infusion and
deliver first of all in to an open system (without pressure
gauge), position of the three-way cocks please see Fig.: 6 - 6.
3. Switch over three-way valve, please see Fig.: 6 - 7, and deliv-
er towards the pressure gauge.
Read off the corresponding value on the pressure gauge and
compare
with the specifications of the TSC for P
max
und P
min
.