System Overview
1
1 - 6 Infusomat® Space 6.0
EN
Opening the Operating Unit Manually
1. Press the locking of the battery compartment cover using a
pointed tool and remove the battery com
partment cover from
the housing.
2. Remove the emergency release crank out o
f the battery com-
partment cover.
3. Press the emergency release cra
nk carefully in the emergency
release plug on the left side of the housing until stop.
The hexagon socket of the emergency release plug is tapered to
the inside. If the emergency release crank is not inserted until
stop, the hexagon socket may be damaged.
Fig.: 1 - 3