Unit Diagnosis / Calibration
2
2 - 6 Infusomat® Space 6.0
EN
54 2230 Drive totphase is defective Software defective or obsolete Update software, change ISP proces-
sor PCB if necessary
55 2231 Temperature sensor defective Software defective or obsolete Update software, change ISP proces-
sor PCB if necessary
56 2236 ... 2255 Internal error
57 2260 ... 2389 TCI error Error in relation to TCI therapy
to
gether with special settings
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaS
eD and inform B. Braun
58 2290 ... 2301 Internal error
59 2320 ... 2328 RTM error Error in relation to Ramp and
T
apper therapy together with
special settings
Restart the device, change settings
slightly if necessary
. If the error per-
sists: please load the history file via
HiBaS
eD and inform B. Braun
60 2340 ... 2344 INTBOL error Error in relation to Interval Bolus
therapy together with
special
settings
Restart the device, change settings
slightly if necessary
. If the error per-
sists: Please load the history file via
HiBaS
eD and inform B. Braun
61 2360 ... 2363 PGM error Error in relation to Program
Mode therapy together
with spe-
cial settings
Restart the device, change settings
slightly if necessary
. If the error per-
sists: Please load the history file via
HiBaS
eD and inform B. Braun
62 2380 Internal error
63 2400 ... 2401 TOM error Error in relation to Take over
Mode to
gether with special set-
tings
Restart the device, change settings
slightly if necessary
. If the error per-
sists: Please load the history file via
HiBaS
eD and inform B. Braun
64 2420 Internal error
65 8000 ... 8049 TCI error Error in relation to TCI therapy
to
gether with special settings
Restart the device, change settings
slightly if necessary. If the error per-
sists: Please load the history file via
HiBaS
eD and inform B. Braun
66 8100 ... 8102 TOM error Error in relation to Take over
Mode to
gether with special set-
tings
Restart the device, change settings
slightly if necessary
. If the error per-
sists: Please load the history file via
HiBaS
eD and inform B. Braun
67 9000 ... 9999 Internal error
Error Code Definition Possible Cause Fault Clearance
Table 2 - 3 Error Codes of the Control Microprocessor (Part 3 of 3)