Disassembly / Assembly
3 - 28 Infusomat® Space 6.0
EN
Injuries through hot soldering iron tip and development of dan-
gerous vapours!
RISK OF INJURY AND POISONING!
- Have soldering work carried out only by trained and author-
ised persons and in suitable premises.
9. Unsolder the two connection leads (Fig.: 3 - 26 / Item 2) at
the pump drive motor (Fig.: 3 - 26 / Item 3).
10. Pull the linear potentiometer carefully out of
the lock bolt
drive.
The linear potentiometer is held in the lock bolt drive by two top
snap-in pins.
Fig.: 3 - 26
Legend of fig. 3 - 26:
Item Designation
1 Locking opening for snap-in pins
2 Connection leads to the pump drive motor
3 Pump drive motor