Procedural Instructions on the TSC
8
8- 4 Perfusor® compact, 3.0 gb
11. Wait until the current step gauge is completely released. Then
remove current step gauge and close syringe holder slowly.
Syringes
1. Check syringe selection (see “Syringe / Syringe Selection”
p. 3 - 14).
Note
internal = table in ROM
EEPROM = free type
internal and EEPROM = table and free type
2. Check whether syringe table is readable.
3. Check syringe recognition.
a) Insert 20 ml potentiometer calibration gauges.
The 20 ml syringe symbol is displayed.
b) Remove potentiometer calibration gauge.
c) Insert max. 50 ml potentiometer calibration gauges.
The 50 ml syringe symbol is displayed.
d) Remove potentiometer calibration gauge.
Electrical Safety
1. Measure the line voltage and note down.
2. Measure protective conductor resistance and note down.
3. Measure patient leakage current as described hereafter and
note down. This is an equivalent leakage current
measurement:
- Remove battery pack or batteries. Unit is at rest.
- Apply nominal voltage + 10 %.
- Measure patient leakage current between short-
circuited mains inlet and plus pole (right top battery
compartment).
- Enter value in check list.
Accessories
Enter accessories, e.g. MFC interface lead or battery in TSC.
Fig.: 8 - 3