15
Technical data
Type 8202/8222
6.3.4. Data of connectors and cables
Number of fixed
connectors
Type of connector
1 male M12 fixed
connector
5-pin female M12 connector (not supplied).
For the M12 connector with order code
917116, use a shielded cable:
• diameter: 3 to 6.5 mm
• wire cross section: max. 0.75 mm
2
1 male M12 fixed
connector and 1
female M12 fixed
connector
5-pin female M12 connector (not sup-
plied) and 5-pin male M12 connector (not
supplied).
For the M12 connector with order code
917116, use a shielded cable:
• diameter: 3 to 6.5 mm
• wire cross section: max. 0.75 mm
2
6.3.5. Sensor specifications of the 8222
Conductivity sensor C=0.01
• Measurement range
• Type of fluid
• 0.05 µS/cm to 20 µS/cm
• ultra-pure water, pure water
Conductivity sensor C=0.1
• Measurement range
• Type of fluid
• 0.5 µS/cm to 200 µS/cm
• pure water, industrial
wastewater
Conductivity sensor C=1
• Measurement range
• Type of fluid
• 5 µS/cm to 10 mS/cm
• industrial wastewater,
wastewater
6.3.6. pH/Redox probe of the 8202
The pH or redox probe must satisfies the following specifications:
• combined probe;
• length: 120 mm;
• with PG 13.5 head;
• with an S7/S8 fixed connector;
• without temperature probe.
The specifications of the probe can be found in the manual
of the probe used.
English