65
Maintenance and troubleshooting
Red
LED
Current
output
Transistor
output
Icon Message
displayed in the
Info menu
Possible cause Recommended action
OFF 4...20 mA depending
on
thresholds
+
"S RTC Reinit" The date and time are
lost because the device
has not been powered
up for at least 5 days.
The message is only
displayed at the first
powering up.
→ Set the date and
time again (see
chpt. 9.11.2).
→ charge the device for
at least 4 hours so that
the date and time are
battery fed for 5 days.
ON 22 mA
1)
depending
on
thresholds
ERR
+
"E:Conductivity" The fluid conductivity is
out of range.
The message is
displayed if the
monitoring of the fluid
conductivity has been
activated, depending on
the set thresholds ERR
LO and ERR HI (see
chpt. 9.13.2).
→ Go into the "Sensor"
function of the
Diagnostic menu to
read the measured
fluid temperature
(chpt. 9.13.2).
→ If necessary, clean and/
or recalibrate the con-
ductivity sensor.
→ If necessary, check the
process.
ON 22 mA
1)
depending
on
thresholds
ERR
+
"E:Polarization" The polarisation slope is
too high.
The message is
displayed if the
monitoring of the
polarisation slope
has been activated,
depending on the set
threshold ERR HI (see
chpt. 9.13.3).
→ Go into the "Sensor"
function of the Diag-
nostic menu to read
the polarisation slope
value (chpt. 9.13.3),
→ If necessary, clean and/
or recalibrate the con-
ductivity sensor.
English
Type 8222 ELEMENT