Section 2: Pump Setup and Programming
Pump Setup &
Programming
26
Reservoir Volume
Enter the volume of uid contained in a lled uid container. e Reser-
voir Volume value decreases as the pump delivers uid or as you prime
the tubing. When you change the uid container, reset the reservoir
volume on this screen. If you do not wish to use the Reservoir Volume
feature, select “Not In Use” (located before 1 and aer 9999 in the range
of values).
e reservoir volume could be set higher than the capacity of the uid
container. Be sure to program the reservoir volume to reect the actual
volume of the medication being used.
Dose Volume
Enter the volume of the dose in milliliters. e maximum volume is
1000 ml. Entering a Dose Volume will automatically reset the cycle and
set the time that the “Next Dose Starts in” to “Immediate.” If you enter a
Dose Volume that would cause the rate to be greater than the maximum
rate of 125 ml/hr, the pump will automatically lengthen the Dose Dura-
tion to accommodate the new volume. You will then need to conrm the
new Dose Duration or enter a dierent duration. e cycle may also be
aected if the Dose Duration is lengthened; in this case, you would also
need to conrm the new Dose Cycle or enter a dierent Dose Cycle.
Dose Duration
Enter the duration for delivery of the dose, up to 24 hours. You will not
be able to select a duration that would cause the rate of delivery of the
programmed Dose Volume to exceed 125 ml/hr. Entering a Dose Dura-
tion will automatically reset the Dose Cycle and set the Dose Starts in to
“Immediate.
Dose Cycle
e cycle is the time from the start of one dose to the start of the next dose.
e programmable values for cycle are based on the Dose Duration. ere
must be at least 5 minutes between the end of one dose and the start of the
next; therefore, the minimum programmable cycle is the Dose Duration
plus 5 minutes. e maximum cycle is 96 hours. Entering a cycle automati-
cally sets the Next Dose Start Time to “Immediate.