Section 2: Pump Setup and Programming
Pump Setup &
Programming
27
KVO Rate
e KVO or Keep Vein Openrate is optional. It allows the delivery of a
medication to help maintain catheter patency. e programming range is 0 to
125 ml/hr. If the time the next Dose Starts in is programmed, the KVO rate is
active during the initial delay. e KVO rate is also active between doses.
Dose Remaining
e Dose Remaining screen is displayed during dosing or if a dose was
interrupted. is screen displays the time in hours and minutes that are
required to infuse the remaining portion of the dose.
NOTE: If the dose is interrupted and the pump is stopped, both the Dose
Remaining and “Dose Starts in screens will be displayed in the programming
screens. When the pump is restarted, the remainder of the interrupted dose
will be delivered. If a full dose needs to be delivered, theDose Starts Inscreen
must be reprogrammed before starting the pump. e user should always
review all of the programming screens to verify when the pump is within the
dosing cycle. eDose Starts inscreen, shown with “— hrs min, allows
the user to reset the time aer which the dose should start.
Dose Starts in
e “Dose Starts in” screen allows you to delay the start of delivery up to
4 days by allowing you to select the amount of time aer which the rst
dose should begin. e pump must be running aer the selected delay in
order for delivery to begin. If a delayed start is not desired, program the
Dose Starts in to “Immediate” so that delivery will begin as soon as you
start the pump.
e delay of the start of the next dose can be programmed in Lock
Level 0 or Lock Level 1 at any time, but if a dose is in progress, this will
cancel the remainder of the dose.
Given
is screen shows the total amount of medication delivered since the last
time this value was cleared. e amount shown is rounded to the near-
est 0.01 ml. If this value reaches 99999.95, it automatically returns to 0
and continues counting. e Given amount does not include medication
delivered with the priming feature.