74
Reference
Section 5: Reference
Cassette Sensor Check
Unlock the cassette by inserting a coin into the lock and turning
clockwise.
e pump will sound a continuous two-tone alarm and the display
should show “No Disposable, Clamp Tubing.
Press
or
to silence the alarm. Press and hold
Å
to turn
the pump o.
e following 3 checks (LCD, Motor and Gear Train and Reservoir Volume
Empty Alarm Check) should be performed in the sequence shown.
LCD Check
With the pump turned o, press
Å
. e second screen that the
pump displays will consist of 32 blocks of characters. Examine the
LCD to verify that there are no missing pixels (dark dots) in the char-
acter blocks.
Motor and Gear Train Check
Program the Reservoir Volume to 2.0 ml.
Attach a 50 or 100 ml Medication Cassette Reservoir or CADD Ad-
ministration Set to the pump. Lock the cassette.
Press and hold
until 3 series of dashes appear. Release
.
Press and hold
. While priming the tubing, listen to the motor
for excessive noise or grinding sounds. Count the number of pump
activations. e pump should prime 10 double activations and then
stop. Press
to return to the main menu.
Reservoir Volume Empty Alarm Check
Program the Reservoir Volume to 1.0 ml. Press
until Reservoir
Volume is displayed on the LCD. Press
´
or
Î
until 1.0 ml is
displayed. en press
¤
.