Cassette Sensor Check
• Unlock the cassette by inserting a coin into the lock and turning
clockwise.
• e pump will sound a continuous two-tone alarm and the display
should show “No Disposable, Clamp Tubing.”
• Press
⁄
or
„
to silence the alarm. Press and hold
Å
to turn
the pump o.
e following 3 checks (LCD, Motor and Gear Train and Reservoir Volume
Empty Alarm Check) should be performed in the sequence shown.
LCD Check
• With the pump turned o, press
Å
. e second screen that the
pump displays will consist of 32 blocks of characters. Examine the
LCD to verify that there are no missing pixels (dark dots) in the char-
acter blocks.
Motor and Gear Train Check
• Program the Reservoir Volume to 2.0 ml.
• Attach a 50 or 100 ml Medication Cassette Reservoir or CADD™ Ad-
ministration Set to the pump. Lock the cassette.
• Press and hold
‹
until 3 series of dashes appear. Release
‹
.
Press and hold
‹
. While priming the tubing, listen to the motor
for excessive noise or grinding sounds. Count the number of pump
activations. e pump should prime 10 double activations and then
stop. Press
„
to return to the main menu.
Reservoir Volume Empty Alarm Check
• Program the Reservoir Volume to 1.0 ml. Press
„
until Reservoir
Volume is displayed on the LCD. Press
´
or
Î
until 1.0 ml is
displayed. en press
¤
.