75
Reference
Section 5: Reference
• Press and hold
‹
until 3 series of dashes appear. Release
‹
.
Press and hold
‹
. e pump should prime 10 double activations
and then stop. e pump will alarm and display “Reservoir Volume
Empty.” Press
„
.
Starting/Stopping the Pump
• Program the pump with the following values:
Reservoir Volume: 1.0 ml
Continuous Rate: 50 ml/hr
Given: 0.00 (press
¤
)
• Program the Air Detector O (see Section 4, Biomed Functions).
• Press and hold
⁄
. “Starting” appears followed by 3 sets of dashes,
each accompanied by a beep. A review of the programmed param-
eters then appears. e main screen should appear with “RUN” in the
display.
• To stop the pump, press and hold
⁄
. “Stopping” appears followed
by 3 sets of dashes that disappear one at a time, each accompanied
by a beep. e main screen should appear with “STOPPED” in the
display.
Activation Timing Check
• Reprogram the Reservoir Volume to 1.0 ml and clear the Given
screen.
• Press and hold
⁄
until 3 dashes disappear from the display. e
pump should sequentially display all of the programmed values. Start
a timer at the rst motor activation.
• Count the activations. One activation should occur every 6 seconds.
Approximately 66 seconds and 10 activations later, the Reservoir Vol-
ume empty alarm should occur. e display should show “Reservoir
Volume empty.”