Section 5: Reference
GIVEN Check
• Stop the pump by pressing and holding
⁄
. Press
„
to advance to
the Given screen. e display should show 1.00 ml. (If the above steps
have not been followed exactly, a dierent value may appear.)
• Press
¤
. e display should now show 0.00 ml.
Air Detector Test
• Turn the Air Detector On (see Section 4, Biomed Functions).
• Reprogram the Reservoir Volume to 10.0 ml. Press „ until Reser-
voir Volume is displayed on the LCD. Press
´
or
Î
until 10.0 ml
is displayed. en press
¤
.
• Attach an empty 50 or 100 ml Medication Cassette Reservoir or
CADD™ Administration Set to the pump. Secure it using the lock
button.
• read the tubing through the Air Detector groove.
• Start the pump.
• e pump should respond with a continuous two-tone alarm and the
display should read “Air In Line Detected.”
• Press
„
or
⁄
to silence the alarm and remove the empty Medi-
cation Cassette Reservoir or CADD™ Administration Set.
• Now attach a Medication Cassette Reservoir containing uid, or a
primed CADD™ Administration Set to the pump. Lock the cassette.
Make certain that there is no air in the uid path.
• read the tubing into the Air Detector groove.
• Start the pump. e pump should deliver without activation of the air
detection alarm.