Upstream Occlusion Sensor Test
• Verify the Upstream Occlusion Sensor is turned On (see Section 4,
Biomed Functions).
• Obtain a CADD™ Administration Set with bag spike. Also obtain a
clamp (slide clamp or hemostat).
• Insert the CADD™ Administration Set spike into an appropriate,
standard IV bag lled with water. Attach the cassette to the pump.
Prime the entire uid path.
• Program the pump to deliver a continuous maximum rate. Press and
hold
⁄
to start the pump.
• Clamp the tubing halfway between the IV bag and the pump. e
pump should alarm within 3 activations aer clamping the tubing.