Section 5: Reference
Upstream Occlusion Sensor Test
Verify the Upstream Occlusion Sensor is turned On (see Section 4,
Biomed Functions).
Obtain a CADD Administration Set with bag spike. Also obtain a
clamp (slide clamp or hemostat).
Insert the CADD Administration Set spike into an appropriate,
standard IV bag lled with water. Attach the cassette to the pump.
Prime the entire uid path.
Program the pump to deliver a continuous maximum rate. Press and
hold
to start the pump.
Clamp the tubing halfway between the IV bag and the pump. e
pump should alarm within 3 activations aer clamping the tubing.