84
Reference
Section 5: Reference
Volumetric Accuracy Testing
Description
A predetermined amount of water is delivered into a collection device
such as a burette or graduated cylinder. e amount of water delivered is
compared to the amount that the pump is programmed to deliver.
Nominal system accuracy is given in the technical specications section
for the pump. at is, under the test conditions described below, the ac-
curacy of the pump and Medication Cassette Reservoir will be nominal
with a 90% condence level. e nominal test conditions are as follows:
degassed water at 25 ± 5°C without back pressure.
Equipment needed
50 or 100 ml Medication Cassette Reservoir with attached CADD
Extension Set with Anti-Siphon Valve OR
50 or 100 ml Medication Cassette Reservoir with ow stop feature
with attached CADD Extension Set (catalog numbers start with
21-73xx)
50 or 60 ml syringe
A uid collection device such as a burette or a Class A, 25 ml capac-
ity graduated cylinder
40 ml of room temperature water
Procedure
1. Fill the 50 or 60 ml syringe with 40 ml of water. Transfer the water
into a Medication Cassette Reservoir.
2. Remove any air from the Medication Cassette Reservoir by aspi-
rating the air with the syringe. Attach the CADD Extension Set.
Prime the tubing so it is lled with uid to the end of the extension
set Luer.
3. Attach the end of the extension set to the uid collection device.
4. Attach the cassette to the pump. Program the Reservoir Volume to
20 ml. is is the intended delivery volume. Remove all clamps.