85
Reference
Section 5: Reference
5. Program a Continuous Rate of 125 ml/hr. Start the pump and wait
for the reservoir volume empty alarm.
6. When delivery is complete, record the volume of uid delivered.
is is the actual delivery volume.
7. Find the dierence between the volume of the amount delivered
and the intended delivery volume. is is the inaccuracy volume.
8. Divide the inaccuracy volume by the intended delivery volume and
multiply by 100. is is the accuracy error percentage.
9. If the accuracy error percentage is greater than ± 6%, repeat the
test with a new Medication Cassette Reservoir. If the pump fails
a second time, call Smiths Medical MD, Inc. or Smiths Medical
International Ltd.
Example: Actual Delivery Volume: 19.5 ml
Intended Delivery Volume: – 20.0 ml
Inaccuracy Volume: = –0.5 ml
Inaccuracy Volume: 0.5 ml
Intended Delivery Volume: ÷ 20.0 ml
Accuracy Error: = –0.025
Accuracy Error: -0.025
× 100.00
Accuracy Error Percentage: = –2.5%