LCR Meter 11021 / 11021-L User’s Manual
2-4
2. The accuracy of inductance, capacitance and resistance
The quality factor is 10 (dissipation factor 0.1), relative to the impedance accuracy,
where
Inductance = 2πfL
Capacitance = 1/2 fC
based on the Figure 2-1 LC and Impedance Conversion Table.
When the quality factor <10 (dissipation factor >0.1), the inductance accuracy times (1+
1/Q) while the capacitance accuracy times (1+D).
3. Dissipation Factor, Quality Factor, AC Impedance Accuracy
Dissipation Factor
Dissipation Factor Accuracy= ±
tane × (1 + D
2
)
1 – D × tane
Quality Factor 10
Quality Factor Accuracy = ±
tane × (1 + Q
2
)
1 – Q × tane
Quality Factor < 10, Quality Factor Accuracy times (1 + 1/Q)
Ú The e is the error spec in Table 2-1.
4.
Equivalent Series Resistance and Equivalent Parallel Resistance Accuracy
Quality Factor 0.l
AC Impedance Accuracy = Impedance Accuracy
Quality Factor 0.l, Accuracy times (l + Q)