Specification
2-3
Table 2-2 shows the basic accuracy of 11021-L:
• The accuracy times 2 when is it is measured in high speed.
10 MΩ 4.00%
0.40°
1MΩ 2.00% 3.00% 5.00% 8.00%
0.25° 1.50° 2.50° 2.50°
330 kΩ 0.20% 0.30% 1.00% 1.00%
33 kΩ 0.12° 0.50° 1.50° 1.50°
33 kΩ 0.10% 0.15% 0.30% 0.40%
3.3 kΩ 0.10° 0.30° 0.30° 0.30°
3.3 kΩ 0.10% 0.15% 0.30% 0.40%
300 Ω 0.10° 0.10° 0.30° 0.30°
300 Ω 0.20% 0.20% 0.30% 0.40%
30 Ω 0.10° 0.12° 0.80° 0.80°
30Ω 0.25% 0.30% 0.40% 0.40%
1Ω 0.15° 0.20° 0.30° 0.40°
1.10% 1.70% 2.00% 2.00%
100 mΩ 0.90° 2.50° 5.00° 6.00°
Set
Range
Range 1kHz 10kHz 40kHz 50kHz
Impedance
Z
Freqency
10 kΩ
1 kΩ
100 Ω
10 Ω
100 kΩ
Table 2-2 1V|Z|,θ Accuracy
10 MΩ 5.00%
3.50 °
1MΩ 2.00% 3.00% 5.00% 8.00%
0.25 ° 1.50 ° 2.50 ° 2.50 °
330 kΩ 0.20% 0.75% 1.00% 1.00%
33 kΩ 0.12 ° 0.50 ° 1.50 ° 1.50 °
33 kΩ 0.10% 0.15% 0.75% 1.00%
3.3 kΩ 0.10 ° 0.30 ° 0.75 ° 0.75 °
3.3 kΩ 0.10% 0.15% 0.75% 1.00%
300 Ω 0.10 ° 0.10 ° 0.75 ° 0.75 °
300 Ω 0.20% 0.20% 0.30% 1.00%
30 Ω 0.10 ° 0.12 ° 0.80 ° 2.00 °
30Ω 0.25% 0.30% 0.40% 1.00%
1Ω 0.15 ° 0.20 ° 0.30 ° 1.00 °
1.10% 1.70% 2.00% 2.00%
100 mΩ 0.90 ° 2.50 ° 5.00 ° 6.00 °
Set Range Range 1kHz 10kHz 40kHz 50kHz
Impedance
Z
Freqency
10 kΩ
1 kΩ
100 Ω
10 Ω
100 kΩ
Table 2-3 50mV|Z|, θ Accuracy