GPIB Commands
5-7
BINNing Command Set
(1) Command: BINNing:UPPer:BIN{1∼8}
Parameter: {?| , <number>}
Return: The high limit of each BIN, the format is < NR3 >.
Function: It sets the high limit of each BIN within the value –9999.99 ~ +9999.99.
To set BIN2 to +99.89%, send the command BINNing:UPPer:BIN2
+99.89; and send command BINNing:UPPer:BIN2? for query.
Syntax: BINN:UPP:BIN1 99.99
Query Syntax: BINN:UPP:BIN1?
Return Data: 9.999000e+01
(2) Command: BINNing:UPPer:AUX
Parameter: {The high limit of secondary parameter (NR3)}
Return: The high limit of secondary parameter, the format is < NR3 >.
Function: It sets the AUX-HI for BINNING that is the high limit for the secondary
measurement parameter (–9.9999E11 ~ +9.9999E11).
Syntax: BINN:UPP:AUX 999.9999E9
Query Syntax: BINN:UPP:AUX?
Return Data: 9.999990E11
(3) Command: BINNing:LOWer:BIN{1∼8}
Parameter: {?| , <number> (NR2 mode)
Return: {The low limit of each BIN}
Function: It sets the low limit of each BIN. The value is –9999.99 ~ +9999.99.
Syntax: BINN:LOW:BIN1 -99.99
Query Syntax: BINN:LOW?BIN1?
Return Data: -9.999000e+01
(4) Command: BINNing:LOWer:AUX
Parameter: {The low limit of secondary parameter (NR3)}
Return: {The low limit of secondary parameter (NR3)}
Function: It sets the AUX-LO for BINNING that is the low limit of the secondary
measurement parameter (–9.9999E11 ~ +9.9999E11).
Syntax: BINN:LOW:AUX -999.9999E9
Query Syntax: BINN:LOW:AUX?
Return Data: -9.9999990E11
(5) Command: BINNing:NOMinal
Parameter: {The nominal of primary parameter (NR3)}
Return: {The nominal of primary parameter (NR3)}
Function: It sets the NOMINAL for BINNING that is the nominal of the primary
measurement parameter (0.0000E-16 ~ +9.9999E12).
Syntax: BINN: NOM 0.9999E+9 Set the center value to be
0000.9999G..
Query Syntax: BINN:NOM?
Return Data: 9.9990000E8
(6) Command: BINNing:STATe
Parameter: {ON (1) | OFF (0) }
Return: {1 | 0}
Function: It sets the BINNING to on or off.
Syntax: BINN:STAT ON