LCR Meter 11021 / 11021-L User’s Manual
5-8
Query Syntax: BINN:STAT?
Return Data: 1
(7) Command: BINNing:RESult?
Parameter: None
Return: {The result of BINNING (+0 ∼ +9)}
Function: It queries the BINNING result. If BINNING is OFF or the sorting result is
0, it outputs +0. If the sorting result is OUT, it outputs +9.
Query Syntax: BINN:RES?
Return Data: 1
CALCulate Command Set
(1) Command: CALCulate1:FORMat
Parameter: {REAL|MLINear|CP|CS|LP|LS|ZS|RS|RP}
Return: {Rp|Zs|Cp|Cs|Lp|Ls|Zs|Rs|Rp}
Function: It sets or queries the primary parameter of present measurement.
Description: REAL The real part of impedance
MLINear The absolute of impedance
CP The equivalent parallel capacitance
CS The equivalent serial capacitance
LP The equivalent parallel inductance
LS The equivalent serial inductance
ZS The equivalent serial impedance
RS The equivalent serial resistance
RP The equivalent parallel resistance
Syntax: CALC1:FORM REAL
Query Syntax: CALC1:FORM?
Return Data: Rp (Rs)
PS: Set the status return to be Rp or Rs by the command [:SENSe]:FUNCtion[:ON].
(2) Command: CALCulate2:FORMat
Parameter: {IMAGinary|PHASe|D|Q|REAL|RS|XS}
Return: {Xs|PHAS|D|Q|Rs|RS|Xs}
Function: It sets or queries the secondary parameter of present measurement.
Description: IMAGinary The imaginary part of impedance
PHASe The phase
D The dissipation factor
Q The quality factor
REAL The real part of impedance
RS The equivalent serial resistance
XS The serial equivalent of imaginary part
Syntax: CALC2:FORM Q
Query Syntax: CALC2:FORM?
Return Data: Q
(3) Command: CALCulate{1|2}:LIMit:BEEPer:CONDition
Parameter: {FAIL|PASS}
Return: {FAIL|PASS}
Function: It defines the beeper output.
Description: FAIL The beeper enables when the comparison result is FAIL.
PASS The beeper enables when the comparison result is PASS.
Syntax: CALC1:LIM:BEEP:COND FAIL