GPIB Commands
5-11
2 The DUT has no contact
<DAT1> The test data of primary parameter
<DAT2> The test data of secondary parameter
Following outputs only when the comparator is on:
<CMP1> The compared result of primary parameter
<CMP2> The compared result of secondary parameter
0 No secondary parameter
1 The tested data is within the range
2 The tested data is too high
4 The tested data is too low
8 The DUT has no contact
Following outputs only when BINNING is on:
<BIN> The sorting result.
0 The secondary parameter is Fail
18 The primary and secondary parameters are Pass
9 The primary parameter is Fail
Query Syntax: FETC?
Return Data: 0,1.314969e+00,1.176097e+02,2
PS: <STATE>:0;<DAT1>:1.314969e+00;<DAT2>:1.176097e+02;<BIN>:2
[:SENSe] Command Set
(1) Command: [:SENSe]:CORRection:COLLect[:ACQuire]
Parameter: {STANdard1|STANdard2}
Return: None
Function: It executes OPEN and SHORT correction procedure.
Description: STANdard1: Execute OPEN procedure.
STANdard2: Execute SHORT procedure.
Syntax: CORR:COLL STAN1
(2) Command: [:SENSe]:CORRection:DATA?STANdard
Parameter: {STANdard1| STANdard2}
Return: Two correction values, the format is < NR3 >
Function: It queries the corrected value.
STANdard 1 is the corrected value of OPEN that is G and B.
STANdard 2 is the corrected value of SHORT that is R and X.
Query Syntax: CORR:DATA? STAN1
Return Data: 1.451959e-11,5.616254e-09
(3) Command: [:SENSe]:FIMPedance:APERture
Parameter: FAST, MEDium, SLOW
Return: FAST, MEDIUM, SLOW
Function: It sets or queries the measurement speed.
Syntax: FIMP:APER MED
Query Syntax: FIMP:APER?
Return Data: MEDIUM
(4) Command: [:SENSe]:FIMPedance:RANGe:AUTO
Parameter: {ON(1) |OFF(0) }
Return: {1|0}
Function: It sets or queries if auto range selection mode is enabled.