7-21
7.2.8 POWer Sub-system
POWer
:L1
:L2
:RISE
:FALL
POWer
Description Set A or B Static in CP mode for load.
Syntax POWer <NR1 | CHAR>
Parameter 0 | B, 1 | A
Example POW 0 Set CP mode to Static B.
POWer:L1/L2
Description Set the Static Load Current of constant power mode.
Syntax POWer: L1 <NRf+>[suffix]
POWer: L1 <NRf+>[suffix]
Parameter For valid value range refer to respective specification.
Example POW:L1 20 Set Constant Power = 20W for Static Load L1.
POW:L2 10 Set Constant Power = 10W for Static Load L2.
POW:L1 MAX Set Constant Power = max. value for Static Load
L1.
POW:L2 MIN Set Constant Power = min. value for Static Load
L2.
Query Syntax POWer:L1?
POWer:L2?
POWer:L1? MAX
POWer:L2? MIN
Return Parameter <NR2> [Unit=Watt]
Query Example POW:L1? Return the set power of Static Load L1.
Return Example 3.12
POWer:RISE/FALL
Description Set current slew rate of constant power static mode.
Syntax POWer:RISE <NRf+>[suffix]
POWer:FALL <NRf+>[suffix]
Parameter For valid value range refer to respective specification.
Example POW:RISE 2.5 Set rise slew rate to 2.5A/µs for static load.
POW:FALL 1A/µs Set fall slew rate to 1A/µs for static load.
Query Syntax POWer:RISE?
POWer:FALL?
POWer:RISE? MAX
POWer:FALL? MIN
Return Parameter <NR2> [Unit=A/µs]
Query Example POW:RISE? Return the rise slew rate of static load.
Return Example 2.5

Table of Contents

Other manuals for Chroma 63200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200 Series and is the answer not in the manual?

Chroma 63200 Series Specifications

General IconGeneral
BrandChroma
Model63200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals